Esprit Ventilator, Model Number V1000, Power Supply PN 1015852 and Power Supply Field Replacement Unit (FRU) PN, Continu
FDA device recall Z-0047-2007 · posted 2006-10-19 · Terminated
Recall details
- Recalling firm
- Respironics California Inc
- Reason for recall
- This action is being initiated due to power supply snubber board failures on certain Esprit Ventilators in countries that utilize operating voltages of 200-240 Volts AC and will involve the replacement of the Esprit Ventilator power supply snubber board. Returned power supplies have shown evidence of overheating and in some cases in which a fan also failed, have shown signs of fire damage.
- Action taken
- The field action will be conducted through consignees (distributors) in the affected countries. The distributors were notified of the field action by Federal Express au on August 8th, 2006. The distributors have been informed that Respironics will provide replacement circuit boards at no cost and will be instructed to upgrade the ventilator power supplies in their region. Instructions have been provided to the distributors and users on how to evaluate the ventilator to be able to continue to use it pending the completion of the upgrades. The distributors will document the replacement of the affected component by signing and dating the activity on the Ventilator Field Action Log (reference sample on Appendix IV). The serial number of the power supply destroyed will also be recorded on the log.
- Root cause
- Other
- Status
- Terminated
- Product code
- CBK
- Quantity affected
- 950
- Distribution
- Worldwide, including Algeria, Argentina, Australia, Bangladesh, Brazil, Cambodia, Chile, Egypt, France, Hong Kong, India, Italy, Jordan, Lebanon, Malaysia, Pakistan, Peru, Philippines, Russia, S Afric
- Date initiated
- 2006-08-08
- Date posted
- 2006-10-19
- Date terminated
- 2010-01-08
- Location
- Carlsbad, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| ESPRIT VENTILATOR (K981072) | Respironics, Inc. | 1998-11-12 |