Esprit V1000 and V200 Ventilators, Model No. V1000 and V200; and the 3rd Generation Power Supply Repair Part (RP)Kits Mi
FDA device recall Z-0008-2015 · posted 2014-10-21 · Terminated
Recall details
- Recalling firm
- Respironics California Inc
- Reason for recall
- A failure of a specific component may prevent the ventilators from operating on AC power or transitioning back to AC power when operating on battery. Additionally, if a battery is not present or is depleted the ventilators will not operate. This failure will cause the loss of ventilator support, which may result in hypercarbia or hypoxemia.
- Action taken
- Philips Healthcare sent an Urgent - Field Safety Notice dated September 17, 2014, to all customers who purchased the Esprit V1000 and V200 Ventilators to inform them that Respironics is recalling Esprit V1000 and V200 Ventilators installed with 3rd Generation Power Supplies and also 3rd Generation Power Supply Repair Part Kits. The letter identifies the affected devices, informs the customers of the problems, hazard involved, and the actions to be taken. Customers were reminded to be advised of the warnings provided. A Philips Respironics representative will contact customers to schedule a no-cost replacement for further information or support. Customers with questions were instructed to contact Philips Healthcare at 1-800-722-9377. For questions regarding this recall call 760-918-7300.
- Root cause
- Unknown/Undetermined by firm
- Status
- Terminated
- Product code
- CBK
- Quantity affected
- 2,206 units total (665 units in US)
- Distribution
- Worldwide Distribution - USA, Argentina, China, Egypt, Ethiopia, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Jordan, Libyan Arab Yamahiriya, Malaysia, Mali, Mexico, Myanmar (Burma). Ne
- Date initiated
- 2014-09-17
- Date posted
- 2014-10-21
- Date terminated
- 2016-09-28
- Location
- Carlsbad, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| ESPRIT VENTILATOR, MODEL V1000 (K034040) | Respironics, Inc. | 2004-01-21 |
| ESPRIT VENTILATOR WITH RESPIRATORY MECHANICS (K023350) | Respironics California, Inc. | 2002-11-06 |
| ESPIRT (K001208) | Respironics, Inc. | 2000-05-12 |
| ESPRIT VENTILATOR (K981072) | Respironics, Inc. | 1998-11-12 |