Atlas FDA

ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222

FDA device recall Z-2064-2026 · posted 2026-05-01 · Open, Classified

Recall details

Recalling firm
Erbe Medical, LLC
Reason for recall
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
Action taken
On 3/31/2026, correction notices were emailed to Medical Directors, Risk Managers, Medical Device Safety Officers, or Healthcare Professionals informing them to take the following actions: 1) Discontinue use of unsupported third-party Scope Port Connectors. 2) Provide this notice to all personnel within your organization who use, service, purchase, stock, or otherwise manage. 3) If you further distributed any affected devices, promptly forward this notice to those accounts and request that they take the same actions. 4) Continue following the instructions for use, including verifying connector compatibility, ensuring that all connectors are fully seated, confirming that all joints and interfaces are secure, and verifying correct placement and alignment of the tubing within the pump head. 5) Utilize the compatibility table in the notice until updated Notes on Use is received. 6) Complete and return the response form via email to fieldaction.medical.us@erbegroup.com. Firm contacts: Correction: fieldaction.medical.us@erbegroup.com, 480-564-1899. Routine Customer Service and Complaints: techservices@erbegroup.com, 1-800-778-3723
Root cause
Component change control
Status
Open, Classified
Product code
OCX
Quantity affected
5,500
Distribution
US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS
Date initiated
2026-03-31
Date posted
2026-05-01
Location
Tempe, AZ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ERBEFLO 2 Disposable Tubing System (K190469)Erbe USA, Inc.2019-05-23