Enzymatic Creatinine (ECREA) Flex(R) reagent cartridges, K1270 for the Dimension Vista(R) 500 System The ECREA method is

FDA device recall Z-2154-2011 · posted 2011-05-09 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Results may be falsely depressed up to 0.4 mg/dL or falsely elevated up to 0.6 mg/dL across the assay range.
Action taken
Siemens sent an Urgent Field Safety Notice letter dated June 2010 to all Dimension Vista(R) ECREA customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to: Discontinue use of ECREA method and transition to the Dimension(R) Jaffe creatinine (CREA) method. To notify anyone to whom they may have distributed the product. Complete and return the Effectiveness Check Questionnaire attached to the letter by fax to (302) 631-8467. For questions regarding this recall call (302) 631-6299.
Root cause
Device Design
Status
Terminated
Product code
JFY
Quantity affected
1,399
Distribution
Nationwide Distribution - including AL, AZ, CA, CO, DE, FL, GA, ID, IL, IA, MD, MA, MI, MT, NH, NJ, NY, NC, OH, OR, PA, SC, TX, VA, WA, and WI.
Date initiated
2010-06-10
Date posted
2011-05-09
Date terminated
2013-12-19
Location
Newark, DE