Atlas FDA

Entroy Pool Lift; a patient lift used in conjunction with a chair or stretcher to lift them in/out of a swimming/therapy

FDA device recall Z-3070-2011 · posted 2011-08-24 · Terminated

Recall details

Recalling firm
Arjo, Inc. dba ArjoHuntleigh
Reason for recall
The Entroy Pool Lift boom may rapidly lower un-commanded to the bottom of its stroke, with a potential for injury to the resident or caregiver.
Action taken
ArjoHuntleigh sent an Urgent Field Correction Notice I.D. 110701AHE to the affected end user via UPS 2nd Day on August 4, 2011. The notice informed the account of the potential for lift boom to rapidly lower un-commanded to the bottom of its stroke. The account was requested to immediately cease use of the affected lift, and disconnect the power to the lift to prevent further use until corrected. The account was requested to complete and sign the Customer Response Form, and return it to ArjoHuntleigh to schedule a service call to correct the lift. Questions were directed to the ArjoHuntleigh at 1-800-323-1245, ext 54485.
Root cause
Device Design
Status
Terminated
Product code
FSA
Quantity affected
1 unit in the U.S., 311 units in the rest of the world
Distribution
Worldwide Distribution - USA including CT
Date initiated
2011-08-04
Date posted
2011-08-24
Date terminated
2011-11-28
Location
Addison, IL

Monitor recalls programmatically