Atlas FDA

Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medical

FDA device recall Z-0055-2025 · posted 2024-10-10 · Open, Classified

Recall details

Recalling firm
Agfa Healthcare NV
Reason for recall
Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.
Action taken
AGFA issued Urgent: Mandatory Correction Important Information on 9/10/24. Letter states reason for recall, health risk and action to take: AGFA HealthCare will provide the software corrections for the affected versions. A patch is available for each of the following XERO Viewer version streams: - 8.4.x - 8.3.x - 8.2.x - 8.1.4.x - 8.1.2.x For versions of XERO Viewer prior to 8.1.2, the issue can be prevented from happening at your site by: - Using the XERO Viewer Extended Tools component (Xtend), if enabled. - Using the Enterprise Imaging Desktops viewer. - There is an option to disable the features affected by this issue. This can be configured by your administrator or contact AGFA Support. Our Support organization representative will contact all customers on any of these versions and coordinate the correction activity. The patch installation will not require any downtime. We request that you complete and return the attached mandatory feedback form. Please direct questions or concerns to your AGFA HealthCare Customer Support, and please reference PRB0762354. AGFA HealthCare Customer Support: 1.866.599.3868
Root cause
Software design
Status
Open, Classified
Product code
LLZ
Quantity affected
801 units
Distribution
US Nationwide distribution. Foreign: Argentina, Australia, Belgium, Chile, Canada, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Georgia, Gr
Date initiated
2024-09-06
Date posted
2024-10-10
Location
Mortsel, Belgium