ENDOPATH ETS Compact Flex (240 mm shaft length) Linear Cutters, Model Number(s): SCW45; CTS45; SCB45; CTS45NK; 6CB45 and
FDA device recall Z-3187-2011 · posted 2011-09-12 · Terminated
Recall details
- Recalling firm
- Ethicon Endo-Surgery Inc
- Reason for recall
- Ethicon Endo-Surgery initiated a voluntary global recall for specific production lots of ENDOPATH ETS Compact Flex 45mm articulating linear cutters because the possibility exists that damage to the blister pack may have compromised the sterility of the device.
- Action taken
- Ethicon Endo-Surgery Inc sent an URGENT DEVICE RECALL letter dated August 11, 2011 to all affected customers. The letter identified the product, the problem, the action to be taken by the customer. Customers were instructed to examine their inventory immediately to determine if they have any of the affected product. Fill out the Business Reply Form and fax it to 1-866-853-2802 if they have any of the affected product. To ensure adequate identification of the return product at their facility customers should complete the Customer Purchase Order Field on the Business Reply Form. For any questions regarding this recall call 1-866-708-2330 and reference Event 2471. For clinical or product support call 1-800-USE-ENDO, option 6, or contact your local sales representative.
- Root cause
- Packaging
- Status
- Terminated
- Product code
- GDW
- Quantity affected
- 27,582 instruments
- Distribution
- Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI an
- Date initiated
- 2011-07-29
- Date posted
- 2011-09-12
- Date terminated
- 2013-02-15
- Location
- Cincinnati, OH
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| ENDOPATH LINEAR CUTTERS AND STAPLERS AND ECHELON60 ENDOSCOPIC LINEAR CUTTERS AND STAPLERS (AND RELOADS) (K070887) | Ethicon Endo-Surgery, LLC | 2007-05-25 |
| ENDOPATH ENDOCUTTER 60 ENDOSCOPIC LINEAR CUTTER, MODELS EC60, SC60 AND LONG60 AND ENDOPATH & PROXIMATE LINEAR CUTTERS (K061156) | Ethicon Endo-Surgery, Inc. | 2006-07-03 |
| PROXIMATE AND ENDOPATH LINEAR CUTTERS AND STAPLERS (K020779) | Ethicon Endo-Surgery, Inc. | 2002-06-05 |