Endo TCS Driver Straight Hexalobe #6 P/N 5210-1004, Non-Sterile, Manufactured by Titan Spine, LLC. Endo TCS driver is on

FDA device recall Z-1324-2015 · posted 2015-03-26 · Terminated

Recall details

Recalling firm
TITAN SPINE, LLC
Reason for recall
The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that involved patients with dense bones. As a result of this malfunction, Titan Spine, LLC is removing all of the TCS Straight Hex lobular #6 Drivers (5210-1004) Lot# G150101 from the field.
Action taken
Titan Spine sent an "Urgent Medical Device Recall" letter via Fed Ex dated February 20, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A replacement TCS Straight Hex lobular# 6 Drivers (5210-1004) will be shipped out to each Titan Spine Area Vice President and Sales Rep. within 4-6 weeks. For further questions please call (262) 242-7801.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
OVE
Quantity affected
26
Distribution
US Distribution to the states of : AZ, CA, CT, FL, GA, ID, IL, KY, MD, MO, NV, NY, OH, OK, OR, TX, UT and WI.
Date initiated
2015-02-20
Date posted
2015-03-26
Date terminated
2015-04-17
Location
Mequon, WI