ENDO KIT, Medline Kit SKU DYKE1577A, Component Number 57469, 10 per case; Medline Convenience Kits containing Single Use

FDA device recall Z-0558-2025 · posted 2024-11-26 · Open, Classified

Recall details

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Reason for recall
Identified lots of Olympus product may be missing sterile and manufacturing lot numbers and expiration dates on the sterile packaging. The information can be found on the outer box or zipper bag label. Medline is recalling its kits which utilize affected product as a component.
Action taken
On October 25, 2024, the firm notified customers of the recall via letters titled "Olympus America, Inc. IMMEDIATE ACTION REQUIRED RECALL". Customers were informed that the specified lot of kits may contain the affected lot of Single Use Biopsy Valve MAJ-1555 recalled by Olympus. The missing sterile and manufacturing lot numbers and expiration dates can be found on the outer box or zipper bag label. Affected product should not be used if you are unable to determine the expiration date. Customers should locate and quarantine all product from affected lot. Customers should complete the response form provided by the firm. Upon receipt of the completed form, the firm will provide over-labels to place on affected inventory, with instructions for staff to remove the affected component prior to using the kit. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Root cause
Labeling Change Control
Status
Open, Classified
Product code
FDE
Quantity affected
100
Distribution
US Nationwide distribution in the state of NY.
Date initiated
2024-10-25
Date posted
2024-11-26
Location
Northfield, IL