Endo GIA 45mm Curved Tip Articulating Vascular/Medium Reload with Tri-Staple Technology, Item Code EGIA45CTAVM
FDA device recall Z-2512-2018 · posted 2018-07-19 · Open, Classified
Recall details
- Recalling firm
- COVIDIEN MEDTRONIC
- Reason for recall
- Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Action taken
- On May 22, 2018, all U.S. consignees received an UPDATED URGENT MEDICAL DEVICE RECALL and RETURN FORM notice via Federal Express or certified. The initial notice was sent to impacted customers on April 26, 2018. Customers are advised to take the following required actions: 1. Immediately discontinue use, and quarantine affected products listed 2. Return affected product as indicated. 3. Complete the Recalled Product Return Form even if you do not have inventory and submit by either method below: * Product purchased directly from Medtronic: feedback.customerservice@Covidien.com or fax to (800) 895-6140. *Product purchased through distributor: FCAMITG@Covidien.com or fax it to (203) 492-7719.
- Root cause
- Process control
- Status
- Open, Classified
- Product code
- GDW
- Distribution
- Nationwide U.S.A. Internationally to: Albania, Algeria, Austria, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Iceland, Ireland,
- Date initiated
- 2018-05-22
- Date posted
- 2018-07-19
- Location
- NORTH HAVEN, CT