ENCORE(R) UNDERGLOVE, Latex, Powder-Free, Smooth Surgical Gloves, REF 2018480, Size 8, Sterile, Manufactured for Ansell
FDA device recall Z-2845-2011 · posted 2011-07-18 · Terminated
Recall details
- Recalling firm
- Ansell Healthcare Products LLC
- Reason for recall
- Gloves do not meet the tensile requirements of ASTM D 3577 reducing the shelf life of the product from 5 years to 2 years.
- Action taken
- Ansell Healthcare Products LLC sent an Urgent Voluntary Recall letter dated June 9, 2011, to all affected customers on June 10 and 13, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disconstinue use and distribution of any of the affected product and return them to the firm. If the customer distributed gloves further, customers were advised to forward a copy of the notification letter along with their tracking number to their customers. Customers were instructed to complete and return the attached Response Form to Ansell by fax at 1-800-722-8155. Returned product would be replaced or credited to the customer.
- Root cause
- Other
- Status
- Terminated
- Product code
- NPD
- Quantity affected
- 271610 pairs
- Distribution
- Nationwide Distribution
- Date initiated
- 2011-06-10
- Date posted
- 2011-07-18
- Date terminated
- 2013-01-10
- Location
- Dothan, AL