ENCORE(R) UNDERGLOVE, Latex, Powder-Free, Smooth Surgical Gloves, REF 2018465, Size 6 1/2, Sterile, Manufactured for Ans

FDA device recall Z-2842-2011 · posted 2011-07-18 · Terminated

Recall details

Recalling firm
Ansell Healthcare Products LLC
Reason for recall
Gloves do not meet the tensile requirements of ASTM D 3577 reducing the shelf life of the product from 5 years to 2 years.
Action taken
Ansell Healthcare Products LLC sent an Urgent Voluntary Recall letter dated June 9, 2011, to all affected customers on June 10 and 13, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disconstinue use and distribution of any of the affected product and return them to the firm. If the customer distributed gloves further, customers were advised to forward a copy of the notification letter along with their tracking number to their customers. Customers were instructed to complete and return the attached Response Form to Ansell by fax at 1-800-722-8155. Returned product would be replaced or credited to the customer.
Root cause
Other
Status
Terminated
Product code
NPD
Quantity affected
176960 pairs
Distribution
Nationwide Distribution
Date initiated
2011-06-10
Date posted
2011-07-18
Date terminated
2013-01-10
Location
Dothan, AL