FDA device recall Z-0159-2024 · posted 2023-10-24 · Open, Classified

Recall details

Recalling firm
EMPATICA SRL
Reason for recall
May have a battery life lasting less than the full 48 hours, requiring the device to be charged more frequently.
Action taken
Empatica S.r.l issued Field Safety Notice via email on 7/14/23. Letter states reason for recall, health risk and action to take: You can check your device lot number by: " From the Alert App On the home page of your MyEmbrace account on the right hand side of the screen you will see your Embrace Serial Number. The lot number is represented by the 5th, 6th and 7th digit of the serial number. " On the packaging of your device On the box that contained your Embrace2 there is an identification label (See picture below). The device Lot number is identified with the LOT symbol. In the picture below the device Lot number is F11. Empatica recommends that you pay increased attention to the product battery level notifications intended to inform you about the remaining battery charge. As noted in the user manual, the Embrace device is not intended to be used as a stand-alone monitoring device. As the system is not meant to substitute for ongoing seizure monitoring practices, patients are strongly advised to ensure that alternative methods of obtaining medical assistance are available at all times, including during device charging sessions Patients are also encouraged to contact Empatica customer support for any concern related to battery life impacting usability of the Embrace device. Customers contacted by the Empatica customer support team regarding battery duration are offered a free-of-charge replacement device.
Root cause
Device Design
Status
Open, Classified
Product code
POS
Quantity affected
1360 units
Distribution
Worldwide distribution - US Nationwide and the countries of Italy, Ireland, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Germany, Finland, Netherlands, Poland, Sweden.
Date initiated
2023-07-13
Date posted
2023-10-24
Location
Milan, Italy