Atlas FDA

EliA Sample Diluent, Article number, 83-1023-01

FDA device recall Z-1253-2017 · posted 2017-02-24 · Terminated

Recall details

Recalling firm
Phadia US Inc
Reason for recall
Erroneous EliA test results related to problems with liquid level detection and diluent pipetting in Phadia 250 instruments, and deformed bottles. Investigation showed problems with the liquid level detection in diluent bottle batch JJCA resulting in aspiration of no or too low volumes of liquid from the diluent bottle and from the dilution plate.
Action taken
Phadia AB, part of Thermo Fisher Scientific, is voluntarily recalling EliA Sample Diluent, Article number, 83-1023-01, lots JM3Y, JS0B, JP87, JK5N, and JU9C, written notices were mailed via FedEx on 01/30/2017. There have been reports of instrument malfunctions for Phadia 250 due to deformed bottles for the below mentioned product and lots. The deformation of the bottle can cause an erroneous volume detection by the instrument which will result in insufficient aspiration of sample diluent, thus generating insufficient dilution of patient samples. The deformation is described as a flange or brim, located on the inside lower part of the vial opening. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER: Stop using the affected lots of the above mentioned product lots Please, return or attest to scrapping of any unused product lots to the manufacturer and order replacement products free of charge. Review your records to determine if you have had instrument problems due to deformed bottles. If you find the instrument error message 3-145 RIGHT ARM LIQUID DETECTION BELOW LOWER LIMIT Target 2, please collect all log files available and/or contact customer support, who will further assist in assessing possible impact on the test results in scope. Fill in the Medical Device Recall return response on page 3 and return to the manufacturer by e-mail. If customers have any questions, please contact Phadia US, Inc. at 1-800-346- 4364, option #2.
Root cause
Under Investigation by firm
Status
Terminated
Product code
NHX
Quantity affected
5957
Distribution
Domestic: AR, CA, CO, FL, GA, IL, IN, KY, LA, MI, MN, MO, MT, NC, NH, NJ, NM, NY, OH, OR, PA, SD, TN, TX, WI, WV Foreign: None VA/DOD: None
Date initiated
2017-01-30
Date posted
2017-02-24
Date terminated
2017-08-07
Location
Portage, MI