EliA dsDNA Well - Product Usage: intended for the In-Vitro quantitative measurement of IgG antibodies directed to dsDNA
FDA device recall Z-1107-2021 · posted 2021-02-11 · Terminated
Recall details
- Recalling firm
- Phadia US Inc
- Reason for recall
- Decreased values for EliA ANA Positive Control when using the EliA dsDNA Well lot BFA37/0142 may potentially cause false positive results (increase in patient test results).
- Action taken
- Phadia US Inc. (Initial Importer/Distributor) issued notification to consignees via phone and FedEx overnight delivery Dec. 17-18 . Letter states reason for recall, health risk and action to take: Customers are asked to perform the following steps: 1.Scrap EliA dsDNA Well lot BFA3Y/0142 and order a replacement free of charge 2. Assess test results from EliA dsDNA Well lot BFA3Y/0142 and determine if retesting of samples is warranted in accordance with internal operating procedures. 3. Complete the Acknowledgement Form in the Medical Device Field Safety Notice and return the response to the contact person described. Any questions concerning this issue, contact Phadia US Technical Support at 1-800-346-4364, option #2.
- Root cause
- Under Investigation by firm
- Status
- Terminated
- Product code
- LSW
- Quantity affected
- 365 kits US
- Distribution
- US Nationwide distribution.
- Date initiated
- 2020-12-14
- Date posted
- 2021-02-11
- Date terminated
- 2022-08-04
- Location
- Portage, MI