Atlas FDA

ELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System

FDA device recall Z-1857-2025 · posted 2025-05-28 · Open, Classified

Recall details

Recalling firm
Onkos Surgical, Inc.
Reason for recall
Potential challenge with intraoperative assembly of the external taper of a single midsection with the internal taper of a mating proximal femur.
Action taken
Onkos Medical notified consignees via email on about 09/22/2024. Consignees were instructed to segregate and quarantine any product on hand immediately, complete and return the response form, and arrange for existing implants to be returned.
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
KRO
Quantity affected
24 units
Distribution
US Nationwide distribution in the states of AZ, CA, FL, FL, MD, MO, NC, TN.
Date initiated
2024-09-22
Date posted
2025-05-28
Location
Parsippany, NJ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ELEOS Limb Salvage System (K161520)Onkos Surgical, Inc.2016-10-28