Atlas FDA

Elekta Precise Treatment Table

FDA device recall Z-1238-2015 · posted 2015-03-19 · Terminated

Recall details

Recalling firm
Elekta, Inc.
Reason for recall
It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.
Action taken
Planned Action by Elekta: 1. Notifications have been sent out. The notification informs users of the possibility treatment table position errors might occur if there is a maintenance error when replacing a potentiometer. The notification recommends establishing a procedure to verify the accuracy of the treatment table position as shown by the treatment room monitor which is performed as part of the daily machine checks. An example procedure is provided. 2. Elekta will establish a procedure service engineers will follow after maintenance involving the relevant potentiometers to ensure proper installation. The procedure will be taught using a video which includes a step by step guide to the correct methodology for installation of the potentiometer assembly, a description of the impact for not installing correctly, and a guide to the new verification of correct installation procedure. The video will be distributed through Elektas learning management system (LMS). The LMS maintains a record of service engineers who successfully complete the training. 3. Elekta will implement an engineering change to the potentiometer assembly which applies to the manufacture of all new Precise Treatment Tables. This change will eliminate the potential for incorrect installation and will prevent a loose grub screw from causing backlash. This change is expected to consist of a 'D' shaped bore that cannot spin on the shaft of the potentiometer. All actions associated with this correction, including service engineer training, will be completed by June 2015. For further questions please call (770) 300-9725.
Root cause
Radiation Control for Health and Safety Act
Status
Terminated
Product code
JAI
Quantity affected
529 US 2951 Worldwide
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, B
Date initiated
2015-01-29
Date posted
2015-03-19
Date terminated
2021-11-05
Location
Atlanta, GA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ELEKTA ONCOLOGY SYSTEMS PRECISE TREATMENT TABLE (K983678)Elekta Instruments, Inc.1999-01-15