Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, containing 6 inner

FDA device recall Z-0227-2025 · posted 2024-11-06 · Open, Classified

Recall details

Recalling firm
Elekta Instrument AB
Reason for recall
Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles.
Action taken
Elekta issued an URGENT IMPORTANT FIELD SAFETY NOTIFICATION to its consignees on 069/23/2023 via email. Then notice explained the problem, clinical impact, and requested the destruction of the affected product. Contact the local Elekta representative for alternative solutions. They also requested that the notice be posted in a place accessible to all users until this action is closed and to notify all appropriate personnel working with this product of the content of the notice.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
HAW
Quantity affected
111 kits (666 needles)
Distribution
US: AZ, TX, FL, CA, NY, GA, VA, NJ, MO, PA, WA,
Date initiated
2024-09-25
Date posted
2024-11-06
Location
Stockholm, Sweden