Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, 2.1mm Diameter / 14 Gauge, containing 6

FDA device recall Z-1397-2024 · posted 2024-04-16 · Open, Classified

Recall details

Recalling firm
Elekta Instrument AB
Reason for recall
Elekta has become aware that Disposable Biopsy Needles (911933) from one batch (837838839) can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles. The sterility of the biopsy needles has not been affected. This issue has been reported from one site.
Action taken
Elekta issued an URGENT RECALL NOTIFICATION to its consignees on 03/05/2024 via email. The notice explained the problem, potential risk, and requested that the consignee advise appropriate personnel and destroy the affected product.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
HAW
Quantity affected
601 needles
Distribution
US and Australia, Austria, Belgium, France, Germany, Greece, Italy, Netherlands, Spain, Sweden, Switzerland
Date initiated
2024-03-15
Date posted
2024-04-16
Location
Stockholm, Sweden