Elecsys Troponin I STAT Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and

FDA device recall Z-1388-2012 · posted 2012-04-11 · Terminated

Recall details

Recalling firm
Roche Diagnostics Operations, Inc.
Reason for recall
An investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of the serum recovery. Correct serum recovery has been confirmed for the affected lots. The issue is caused by the raw material lot (Poly-L-Lysin) instability. Poly-L-Lysin is the
Action taken
The firm, Roche Diagnostics Corporation, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 12, 2012 to all customers that received the affected lots of Elecsys Troponin I or Elecsys Troponin I STAT. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of the affected product; discard the affected product from their inventory according to their site's local regulations; if their facility has distributed the affected product to other sites, ensure that this letter is provided to those sites; complete and return the attached fax form via fax to 1-888-912-8457 and file this letter for future references. If you have any questions about the information contained in this letter, please contact Roche Diagnostics Technical Support, 24 hours a day, seven days a week at 1-800-428-2336.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
MMI
Quantity affected
1,288 kits
Distribution
Worldwide distribution. Argentina, Kenya, Austria, South Korea, Belgium, Kuwait, Bahrain, Morocco, Brazil, Mexico, Switzerland, Nigeria, Chile, New Zealand, Columbia, Philippines, Germany, Pakistan, A
Date initiated
2012-03-12
Date posted
2012-04-11
Date terminated
2013-02-06
Location
Indianapolis, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ELECSYS TROPONIN I AND TROPONIN I STAT TEST SYSTEMS (K082699)Roche Diagnostics2009-08-19