Elecsys 2010 Analyzer Software version 06-07 The Elecsys 2010 analyzer is intended for the in vitro quantitative and qua

FDA device recall Z-1218-2012 · posted 2012-03-16 · Terminated

Recall details

Recalling firm
Roche Diagnostics Operations, Inc.
Reason for recall
Elecsys 2010 customers have reported getting no value and flag 48 results below expected value range after performing automatic dilution for a sample that previously recovered above the measuring range. Incorrect but believable results may occur for the CEA, Rubella IgG or SHBG assays. The other affected assays will generate either no value or a negative result, both of which are not bel
Action taken
Roche sent an Urgent Medical Device Correction letter dated January 31, 2012, via UPS Ground (receipt required) to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to do the following: use one of the workarounds outlined in this Urgent Medical Device Correction (UMDC) if you are running an automatic dilution on an Elecsys 2010 analyzer with software version 06-07; complete the attached faxback form (4600-00-0112) and fax it to 1-888-912-8460; and file this UMDC for future reference. For questions regarding this recall call 1-800-428-2336.
Root cause
Software design
Status
Terminated
Product code
JJE
Quantity affected
Software version 06-07 was installed on Elecsys 2010 analyzers at 13 accounts.
Distribution
Worldwide Distribution - USA including NY, GA, DE, SD, WI, ME, OR, IA, AND CA and the countries of Austria, Cyprus, Czech Republic, Estonia, France, Germany,Greece, Hungary, Iceland, Italy, Latvia, No
Date initiated
2012-01-31
Date posted
2012-03-16
Date terminated
2012-09-06
Location
Indianapolis, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ELECSYS 2010 ANALYZER (K961481)Boehringer Mannheim Corp.1996-06-14