EE- Series Video Naso Pharyngo Laryngo Scope, VNL-1570STK. Intended to provide optical visualization (via a video monito

FDA device recall Z-0820-2012 · posted 2012-01-20 · Terminated

Recall details

Recalling firm
Pentax Medical Company
Reason for recall
The wrong silicone sealant was used and excess sealant was neglected to be removed during repair of the bronchoscope.
Action taken
Pentax sent an "Urgent: Device Recall - Stop Use Immediately" letter/return response form to their customers on 9/30/2011. The letter identified the affected product and asked customers to immediately discontinue the use of their endoscope(s). Customers were to contact PENTAX at 1-800-736-8293 to schedule the return of the affected product(s). The reason for the recall was also explained, and the letter stated that there will be no charge for shipping, evaluation, or repair of the products. Customers can request loaner instruments, if necessary.
Root cause
Employee error
Status
Terminated
Product code
EOX
Quantity affected
one unit
Distribution
Nationwide Distribution.
Date initiated
2011-09-30
Date posted
2012-01-20
Date terminated
2012-11-27
Location
Montvale, NJ