Atlas FDA

Edwards Lifesciences, Miller Balloon Atrioseptostomy Catheter, Sterile EO, Rx Only, REF: 830515F Product Usage: The Mill

FDA device recall Z-1117-2019 · posted 2019-04-25 · Terminated

Recall details

Recalling firm
Edwards Lifesciences, LLC
Reason for recall
Difficulty with atrioseptostomy catheter balloon deflation after deployment, and reports of balloon fragmentation or detachment.
Action taken
On 03/28/19, Urgent Product Recall - Action Required Field Corrective Action notices were mailed FedEx Second Day to customers. The letter identified the affected product problem and actions to be taken. The firm is requesting the return of any existing inventory of these models. They are conducting an investigation of the issues. At this time, they do not have an estimated date as to when replacement product will become available. The firm urges customers to inquire as to the availability of similar products. Customers with additional questions are encouraged to call 800-424-3278 Option 1. Firm requests that customers fill out and return Customer Acknowledgement forms within 10 days of receipt of this notification.
Root cause
Process design
Status
Terminated
Product code
DXF
Quantity affected
1564
Distribution
Worldwide Distribution - US Nationwide in the states of : IA, FL, MD, CA, NY, TX, CT, NC, OR, LA, UT,AK, WA, AZ, SD,CO,IN, KY, WV, TN, MI, AL, MA, VA, PA, MO, IL, GA, PR, MN, AR, OK, WI. Foreign (OUS)
Date initiated
2019-03-28
Date posted
2019-04-25
Date terminated
2023-11-01
Location
Irvine, CA

Monitor recalls programmatically