Edwards Lifesciences Fem-Flex Femoral Arterial Cannula with Duraflo Coating, DIITFA02225, Sterile, RX only For Single Us
FDA device recall Z-0134-2014 · posted 2013-11-01 · Terminated
Recall details
- Recalling firm
- Edwards Lifesciences, LLC
- Reason for recall
- Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation.
- Action taken
- Edwards sent an Urgent Field Safety Notice Product Recall letter dated August 26, 2013 to all affected customer, via Fed-Ex. The product is sold directly to hospitals and distributors. The distributors involved will communicate with their customers directly. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review their entire inventory, quarantine and return the affected products. An acknowledgement form is included for completion to assist in the assessment of the inventory. For questions contact Edwards Customer Service at 800-424-3278.
- Root cause
- Process control
- Status
- Terminated
- Product code
- DWF
- Quantity affected
- 80 units
- Distribution
- Nationwide Distribution.
- Date initiated
- 2013-08-26
- Date posted
- 2013-11-01
- Date terminated
- 2017-02-22
- Location
- Draper, UT