Edwards Lifesciences Crimpers, Model 9100CR23. The Crimpers are used to crimp the transcatheter heart valve onto the del
FDA device recall Z-1134-2013 · posted 2013-04-17 · Terminated
Recall details
- Recalling firm
- Edwards Lifesciences, LLC
- Reason for recall
- Holes in the sterile barrier pouch may be caused by handling of the device during manufacturing or once pouches have been removed from the unit box and stored on the shelf by customers.
- Action taken
- Edwards Lifesciences notified customers via letter and instructed them to maintain the device in its protective unit carton until ready for use and to inspect the pouch for holes or tears prior to use as instructed in the product's labeling. If any holes or tears are found, customers are instructed to contact Edwards Customer Service to arrange for return and replacement of damaged product.
- Root cause
- Package design/selection
- Status
- Terminated
- Product code
- NPT
- Quantity affected
- 5,472 units
- Distribution
- Nationwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK,
- Date initiated
- 2013-03-26
- Date posted
- 2013-04-17
- Date terminated
- 2013-09-16
- Location
- Draper, UT