Eclipse Treatment Planning System; Release 6.5 through 10 (build 7.3.10 through 10.0.34) Varian Medical Systems, Inc. Us

FDA device recall Z-2780-2011 · posted 2011-07-11 · Terminated

Recall details

Recalling firm
Varian Medical Systems, Inc. Oncology Systems
Reason for recall
In affected Eclipse versions, user editing of the MLC Leaf positions may not automatically invalidate calculated dose distributions.
Action taken
Varian Medical Systems, Inc sent a "Urgent Medical Device Correction Urgent Field Safety Notice" letter dated May 13, 2011 to all affected customers. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter included the description of the product, problem and recommendations and includes the release of a service pack in the future. For additional information please contact the Varian Oncology Help Desk at 888-VARIAN5 (888- 827-4265)
Root cause
Software design
Status
Terminated
Product code
MUJ
Quantity affected
10,097***12-12-12 AMENDED TO: 10664 units***
Distribution
Worldwide distribution: Nationwide (USA) and the country of Canada.
Date initiated
2011-05-13
Date posted
2011-07-11
Date terminated
2014-08-15
Location
Palo Alto, CA