Eclipse Treatment Planning System, Client software versions 8.0 through 8.6; Model Number: H48; Product is manufactured
FDA device recall Z-2220-2011 · posted 2011-05-19 · Terminated
Recall details
- Recalling firm
- Varian Medical Systems, Inc. Oncology Systems
- Reason for recall
- An anomaly in Eclipse Treatment Planning System, versions 8.0 to 8.6, whereby Actual SSD (source surface distance) value may not correspond to the Planned SSD value after Planning Approval.
- Action taken
- The firm, Varian Medical, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 4, 2011, to its customers. ***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1) Set the Planning Approved status in Eclipse (Eclipse displays both the Actual and Planned SSD values during plan approval) and verify both SSD values prior to approving the plan. 2)Verify the SSD values in the Eclipse Treatment Plan Report prior to approving the plan for treatment or proceeding with treatment delivery (this report lists both the Actual and Planned SSD values). 3) Prior to approving the plan in RT Chart, verify both the Actual and Planned SSD values in the Parameters Workspace by selecting Field Parameters>Geometry tab. 4) Prior to treatment, they should verify intended SSD values during independent MU calculation. 5)Re-familiarize themselves with the concepts of Actual SSD and Planned SSD by reviewing the listed information. Please advise the appropriate personnel working in your radiotherapy department of the content of this letter. Varian mitigated the product behavior in version 8.9. If you require further clarification, please feel free to contact the Varian Oncology Help Desk ; phone: USA and Canada - 1888-827-4265; Europe: +41 41 749 8844; email: North America: support-americas@varian.com; Australia/New Zealand: support-anz@varian.com; Europe: support-emea@varian.com; South East Asia: seasia.apps.helpdesk@varian.com; China/Asia: china.apps.helpdesk@varian.com; Japan: Japan.Apps.Helpdesk@varian.com; Latin America: soporte.al@varian.com; Internet: Oncology Systems-www.myvarian.com; Varian Medical Systems -www.varian.com; or your local Varian Customer Support District or Regional Manager.
- Root cause
- Software design
- Status
- Terminated
- Product code
- MUJ
- Quantity affected
- 5288 units
- Distribution
- Worldwide distribution.
- Date initiated
- 2011-04-04
- Date posted
- 2011-05-19
- Date terminated
- 2012-06-27
- Location
- Palo Alto, CA