ECG Cable Part Number: 01.57.471017 for use with the SE-1 series, SE-3 series and SE-601 series Smart Electrocardiograph
FDA device recall Z-0041-2012 · posted 2011-10-12 · Terminated
Recall details
- Recalling firm
- Edan Instruments, Inc.
- Reason for recall
- The 3 leads of the cable, RL, LA and LL were labeled in wrong sequence. The left arm (LA) is mislabeled as right leg (RL). The left leg (LL) is mislabeled as left arm (LA). The right leg (RL) is mislabeled as left leg (LL). There have been no reports of injury as a result of this issue.
- Action taken
- Edan Instruments, Inc. sent an Urgent - Medical Device Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that wrongly labeled cable may lead to incorrect ECG waveforms during ECG test, customers were reminded to review previous diagnoses made from this cable to mitigate any inappropriate recordings or treatment made. Customers were instructed to replace the affected ECG Cables once they receive the new ECG Cable, and collect all the affected ECG Cables and then destroy and dispose of them according to local regulations. Questions or concerns were directed to the following numbers: Oceania and Asia: +86 (755) 2689 8322 North America, South America: +86 (755) 2689 9914 Europe: +86 (755) 2669 4684
- Root cause
- Vendor change control
- Status
- Terminated
- Product code
- DPS
- Quantity affected
- 615 Ccables
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Dominica, Ecuador, Colombia, Honduras, Peru, Mexico, El Salvador, Guatemala, Venezuela, Uruguay, Chile, Panama, Nicaragua, Ecuador, Austr
- Date initiated
- 2010-09-10
- Date posted
- 2011-10-12
- Date terminated
- 2011-10-13
- Location
- Shenzhen, Guangdong, China