ECAT PET Scanners (mobile systems), distributed by Siemens Medical Solutions ECAT PET Scanners (mobile systems), and Rev

FDA device recall Z-0819-05 · posted 2005-05-11 · Terminated

Recall details

Recalling firm
CTI PET Systems Inc
Reason for recall
Failure to properly secure the Mobile Bed Supports during transit could cause the support screws on the Patient Handling System to become loose and possibly break due to increased vibration during transit.
Action taken
Written notification via certified US Mail was sent out beginning 04/05/2005 to distributors and end users (with whom CTI had a service contract) to advise them of the problem.
Root cause
Other
Status
Terminated
Product code
KPS
Quantity affected
80 units
Distribution
Nationwide, Canada, the Netherlands, Ireland & Denmark
Date initiated
2005-04-05
Date posted
2005-05-11
Date terminated
2009-09-11
Location
Knoxville, TN