EasyLink Informatics System, software version 3.0, Part Number 1000034941, DADE BEHRING INC., Newark, DE.

FDA device recall Z-1174-2007 · posted 2007-08-22 · Terminated

Recall details

Recalling firm
Dade Behring, Inc.
Reason for recall
Sample ID and Patient results could be associated with the wrong patient
Action taken
The recalling firm issued a Field Corrective Action letter dated 7/3/07 to inform their customers of the problem and that a software patch would be sent for them to download. The recalling firm plans to telephone the customers after the patch has been released to verify that it has been installed.
Root cause
Other
Status
Terminated
Product code
JJE
Quantity affected
72 units
Distribution
Worldwide: USA including states of AZ, CA, FL, IL, MA. MI, MO, NC, OH, PA, TX, and WV. The product was also shipped to other Dade facilities in MO, DE, IL, CT, and Belgium.
Date initiated
2007-07-03
Date posted
2007-08-22
Date terminated
2007-11-07
Location
Newark, DE