Easy Moving Motor Driven Mobile X-Ray Unit, Models SM-20HF and SM-30HF; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio

FDA device recall Z-650/51-03 · posted 2003-04-25 · Terminated

Recall details

Recalling firm
Sedecal USA, Inc.
Reason for recall
Incorrect identification labels and failure of the cretification testing resulting in noncompliance with the U.S> Federal performance standard
Action taken
Sedecal USA sent end user notification letters dated 9/26/02 to the consignees having the affected Sedecal Mobile X-Ray Systems, informing them that the units do not meet with the U.S. federal performance standards, that the labels are in Spanish and that the labels may not be visible after system installation. Sedecal schedule an appointment to conduct performance tests on the mobile x-ray systems to verify their compliance with the federal performance standard, submit the data to FDA, and place the English labels on the units. Sedecal will reimburse the owners for any out-of-pocket expenses incurred. Sedecal requested the accounts to call 800-920-9525 if they had any questions, and to inform Sedecal if they sold the unit, destroyed the unit or otherwise no longer had the unit in their possession.
Root cause
Other
Status
Terminated
Product code
IZL
Quantity affected
9 units
Distribution
North Carolina, Louisiana, New York, New Jersey, Missouri, Kentucky, Texas and Germany
Date initiated
2002-09-26
Date posted
2003-04-25
Date terminated
2005-07-29
Location
Arlington Heights, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
EASY MOVING MOBILE X-RAY UNIT (MODEL SM-HF) (K012663)Sedecal USA, Inc.2001-08-31