E-Z Clean Disposable Laparoscopic Electrode Model Number 0100L and 0100LS. Expiration dates of 05/2018 to 07/2018. Steri

FDA device recall Z-2097-2013 · posted 2013-08-27 · Terminated

Recall details

Recalling firm
Megadyne Medical Products, Inc.
Reason for recall
Megadyne Medical Products is recalling certain lots of E-Z Clean, L-Wire laproscopic Electrode Standard Stem and Split Stem due to potential breach of insulation.
Action taken
Megadyne sent an Urgent: Medical Device Recall Field Safety Notice dated August 7, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customer were also notified with follow-up phone calls and email beginning on August 7, 2013. Customers were instructed to examine their stock immediately to determine if you have any of the products from the affected lots on hand. If so, please remove form service/discontinue use and contact Megadyne Customer Service at 1 (801) 576-9669 or (800) 747-6110 (USA) for return authorization and replacement or credit.
Root cause
Process control
Status
Terminated
Product code
GEI
Quantity affected
1816
Distribution
Class II Recall - Worldwide Distribution - US including the states of CA, HI, IL, MN, NC, NJ, OH, PA, TX, UT and VA.
Date initiated
2013-08-07
Date posted
2013-08-27
Date terminated
2014-01-28
Location
Draper, UT