DYONICS POWER Shaver Blades, 4.5mm Long Full Radius, package of 3 Catalog Number: 7205341 Product Usage: Smith & Nephew

FDA device recall Z-2732-2011 · posted 2011-07-01 · Terminated

Recall details

Recalling firm
Smith & Nephew, Inc. Endoscopy Division
Reason for recall
Product has insufficient weld between the shaft tube and tip, could result in the tip separating from the shaft tube during use.
Action taken
Smith & Nephew notified customers by an "URGENT - PRODUCT RECALL 1st NOTIFICATION" letter dated May 6, 2011. The letter describes the product, problem, and actions to be taken. The letter asked customers to call 800-343-5717 (option 3) or send an email to endo.andreturn@smith-nephew.com for a Return Authorization (RA) Number to return product, obtain credit or order replacement blades. Customers are to complete the form and return to the address provided being sure to reference the RA number. For any questions regarding this recall customers were instructed to call 508-337-3731.
Root cause
Process control
Status
Terminated
Product code
HAB
Quantity affected
555 units
Distribution
Worldwide Distribution -- USA (nationwide) and countries of: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Germany, Hong Kong, Italy, Japan, Mexico, Netherlands, Turkey, and United Kin
Date initiated
2011-05-06
Date posted
2011-07-01
Date terminated
2012-05-09
Location
Andover, MA