DynaRad Phantom Portable X-Ray System; Del Medical Imaging Corporation, 11550 West King Street, Franklin Park, IL 60131-

FDA device recall Z-0049-2007 · posted 2006-10-19 · Terminated

Recall details

Recalling firm
Del Medical Systems Group
Reason for recall
The x-ray tube head assembly of the Phantom Portable X-Ray System may detach from the boom arm assembly. The tube head could fall and contact the patient or operator.
Action taken
Del Medical sent Field Correction Recall letters dated 10/3/06 to the direct accounts, informing them of the potential for the shoulder bolt in the yoke assembly to come loose. If the shoulder bolt comes loose, the tube head could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.
Root cause
Other
Status
Terminated
Product code
IZL
Quantity affected
53 units
Distribution
Nationwide and Thailand
Date initiated
2006-10-03
Date posted
2006-10-19
Date terminated
2008-01-24
Location
Franklin Park, IL