Atlas FDA

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

FDA device recall Z-1326-2025 · posted 2025-03-12 · Open, Classified

Recall details

Recalling firm
Beckman Coulter, Inc.
Reason for recall
The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results
Action taken
An IMPORTANT PRODUCT NOTICE letter dated 11/6/24 was sent to customers. Contact your Beckman Coulter representative if your laboratory has experienced wash wheel dispense arm spin mixer motor errors that are not resolved by initializing. See the System Management Chapter section in the DxI 9000 Access Immunoassay Analyzer Instruction for Use (IFU) for steps to initialize the Analyzer. Your Beckman Coulter service representative will contact you to schedule a part replacement that resolves this issue. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. Please complete and return the enclosed Response Form within 10 days so we are assured you have received this important communication. If you have any questions regarding this notice, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
JJE
Quantity affected
245 units
Distribution
US Nationwide. Global Distribution.
Date initiated
2024-11-13
Date posted
2025-03-12
Location
Chaska, MN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Access TSH (3rd IS) Assay, DxI 9000 Access Immunoassay Analyzer (K221225)Beckman Coulter, Inc.2022-11-10