Atlas FDA

Durex Real Feel, synthetic polisoprene male condom, SKU 02340-89456-14 02340-89455-00 02340-89456-14 02340-89456-14 0234

FDA device recall Z-0993-2019 · posted 2019-03-09 · Terminated

Recall details

Recalling firm
Reckitt Benckiser LLC
Reason for recall
Five batches of the product have failed the specification of 1.0kPa for Burst Pressure.
Action taken
On December 19, 2018, the firm issued Urgent Medical Device Recall letters to wholesale customers. The letter advised customers of the product issue, and reminded customers that the product labeling states, "Should a condom break or leak during use, seek medical assistance as soon as possible, at least within 72 hrs." Customers are asked to do the following: 1. Examine inventory in your distribution center immediately to determine if you have the specified batch in your warehouse inventory. 2. Segregate and return the specified batch. 3. A pre-paid return label is provided by the firm for your convenience to return the product. Contact Stericycle Expert Solutions at 888-843-0252 for assistance if you need additional labels or help with coordination of pickup and removal of product from your facility. Contact Information: Phone: 888-843-0252 Fax: 855-880-6687 Email: rbhealth3552@stericycle.com 4. Alternatively, you may arrange shipment of product directly to Stericycle at the following address: Attn: Event 3552 2670 Executive Drive, Suite A Indianapolis, IN 46241 Dock Hours are 8AM - 4:30PM Monday - Friday Please call to schedule a dock appointment 5. Finally, notify Stericycle representative by completing and returning the Business Response Form via email rbhealth3552@stericycle.com that you have: a. Checked your inventory in your distribution centers and confirm you have NO product of the specified batch. b. You have collected all units in your inventory and product has been shipped back in accordance with the instructions from your Stericycle representative. Please direct any questions or concerns with this recall to Stericycle Representative at 877-448-5308 and reference event 3552. For general inquiry not related to the recall, please contact RB Custo er Service Director, Robert Drennan at 201-220-8531.
Root cause
Under Investigation by firm
Status
Terminated
Product code
MOL
Quantity affected
1761044
Distribution
Nationwide domestic distribution. Foreign distribution to Barbados, Jamaica, Aruba, Guyana, Haiti, and St Maarten.
Date initiated
2018-12-19
Date posted
2019-03-09
Date terminated
2021-03-26
Location
Parsippany, NJ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
DUREX SYNTHETIC POLYISOPRENE MALE CONDOM (K072169)Ssl Americas, Inc.2008-06-19