DRIVER STAR #8 CANNULATED DART-FIRE (TM), COMPRESSION SCREW, SIZE: #8, REF: DSDS-0008, 1 EACH, Rx ONLY, NON-STERILE, EC

FDA device recall Z-2470-2010 · posted 2010-09-22 · Terminated

Recall details

Recalling firm
Wright Medical Technology Inc
Reason for recall
After autoclaving, the epoxy identifier (colored) band bled and there is concern that it may flake.
Action taken
The firm initiated their recall the product by letter on 08/26/2010. The letters were addressed to distributors and hospital administrators with a fax back response form. Please contact Customer Service at 800-238-7117, if you have any questions.
Root cause
Process design
Status
Terminated
Product code
HXX
Quantity affected
13 units
Distribution
TX, PA, TN, VA, GA, and Spain
Date initiated
2010-08-26
Date posted
2010-09-22
Date terminated
2012-08-02
Location
Arlington, TN