Draeger Vapor 3000. Unheated, calibrated anesthetic vaporizer.

FDA device recall Z-0934-2026 · posted 2025-12-23 · Open, Classified

Recall details

Recalling firm
Draeger, Inc.
Reason for recall
A certain component of affected devices was not delivered within specification and contained impurities.
Action taken
Consignees were mailed an Urgent Medical Device Recall notification dated November 2025. The notification recommends that consignees do not use affected vaporizers and return the provided reply card to confirm this. A Draeger representative will contact consignees to arrange replacement of affected vaporizers. If affected vaporizers have been placed into clinical use, Draeger asks that consignees clean the breathing system of the anesthesia device according to the IFU and use of inspiratory filters is recommended. Consignees are to share the recall notification with all users of affected devices to ensure awareness and forward the notice if product was further distributed. Questions regarding operation of devices are to be directed to Drager Service Technical Support at 1-800-437-2437 (Option 2, 2, 2). Questions about the recall notice are to be directed to Michael Kelhart via email at mike.kelhart@draeger.com or phone at 267-664-1131.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
CAD
Quantity affected
188 units
Distribution
Worldwide - US Nationwide distribution in the states of AK, CA, FL, IA, KS, MD, ME, MN, NY, OH, PA, RI, SD, TN, TX, UT, W and the countries of Argentina, Australia, Belgium, Brazil, Cambodia, China, C
Date initiated
2025-11-24
Date posted
2025-12-23
Location
Telford, PA