Draeger TI500. Isolette Infant Incubator. Catalog Number: MU20505. Draeger Medical, Inc., Telford, PA 18969

FDA device recall Z-1051-2008 · posted 2008-03-18 · Terminated

Recall details

Recalling firm
Draeger Medical, Inc.
Reason for recall
Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.
Action taken
The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or Anbu.arasu@draeger.com.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
FMZ
Quantity affected
18 units
Distribution
Worldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK.
Date initiated
2008-01-31
Date posted
2008-03-18
Date terminated
2008-05-08
Location
Telford, PA