Draeger Infinity Acute Care System Monitoring Solution

FDA device recall Z-0714-2014 · posted 2014-01-10 · Terminated

Recall details

Recalling firm
Draeger Medical Systems, Inc.
Reason for recall
After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.
Action taken
An Urgent Medical Device Recall Letter, dated December 2013, was sent to Users. Once available Drager will update components of the IACS Monitoring Solution System.
Root cause
Software design
Status
Terminated
Product code
MHX
Quantity affected
439
Distribution
US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
Date initiated
2013-12-10
Date posted
2014-01-10
Date terminated
2015-05-20
Location
Telford, PA