Atlas FDA

DPM 6 Patient Monitor and DPM 7 Patient Monitor. The product affected is the Mindray IBP module, P/N is 6800-30-50485 fo

FDA device recall Z-0837-2015 · posted 2014-12-19 · Terminated

Recall details

Recalling firm
Mindray DS USA, Inc. dba Mindray North America
Reason for recall
Mindray has identified an issue that involves the IBP module used with the DPM 6 and DPM 7 Monitor where the IBP measurement data may be displayed in an incorrect IBP tile on the monitors display.
Action taken
Mindray DS USA, Inc. sent an Urgent Field Corrective Action Letter dated 10/6/2014 to their customers via certified mail with return receipt. The letter identified the affected product, the problem, adverse effects on the patient, and corrective actions. Customers are asked to either verify that the IBP modules are not affected by following the steps provided in the "Procedures for Verifying IBP Channel Connections;" or to discontinue use of the IBP modules until a Mindray Service Representative contacts them and performs the test to verify the function of the IBP modules. Customers are to contact their local Mindray Service Representative at 1-800-288-2121 to inform them of which corrective action will be taken. Questions should be directed to Diane Arpino at 201-995-8407 or d.arpino@mindray.com.
Root cause
Process control
Status
Terminated
Product code
MHX
Quantity affected
455 units
Distribution
Distributed to the states of AL, CA, CO, FL, GA, IA, IL, IN, LA. MA, MO, MS, MT, NC, ND, NY, OH, PA, SC, TX, VA, WA and WV.
Date initiated
2014-10-06
Date posted
2014-12-19
Date terminated
2017-06-06
Location
Mahwah, NJ