Disposable Tips 1000 l with filter Conductive disposable tips for Tecan Genesis Series * Order Number 612513.1 * Tecan S
FDA device recall Z-0326-06 · posted 2005-12-23 · Terminated
Recall details
- Recalling firm
- Tecan U S Inc
- Reason for recall
- Double filtering of 1000 l tips. The defect could result in cross contamination of other samples due to possible dripping of sample while moving over other samples.
- Action taken
- The consignee was notified via fax''d letter on/about 11/08/2005.
- Root cause
- Other
- Status
- Terminated
- Product code
- JQW
- Quantity affected
- 3000 Tips
- Distribution
- IL and Europe
- Date initiated
- 2005-04-11
- Date posted
- 2005-12-23
- Date terminated
- 2006-05-18
- Location
- Research Triangle Park, NC