Disposable Bimanuals Irrigation and Aspiration Handpiece Set: Item #1720 Disposable Bimanuals Irrigation/Aspiration Set
FDA device recall Z-0741-2007 · posted 2007-06-27 · Terminated
Recall details
- Recalling firm
- Oasis Medical Inc
- Reason for recall
- Plastic shavings are coming from the product handle during irrigation. Use of this product may leave plastic shavings in the patient's eyes.
- Action taken
- An April 9, 2007, notification of the recall will be sent to distributors and users. They are requested to notify down to the user (physician) level. Included in the Recall Notification Letter are instructions: "In addition, if you may have further distributed this product, please identify your customers and notify them at once of this product recall. Your notification to your customers may be enhanced by induding a copy of this recall notification letter. This recall should be carried out to the user level. Your assistance is appreciated and necessary to prevent any inconvenience."
- Root cause
- Other
- Status
- Terminated
- Product code
- KYG
- Quantity affected
- 3050 total for all item numbers
- Distribution
- Worldwide, including USA, United Kingdom, Canada, Spain, Turkey, Kuwait, Colombia, Italy, Greece, Australia, Switzerland, South Africa, Germany, Portugal, and France.
- Date initiated
- 2007-04-09
- Date posted
- 2007-06-27
- Date terminated
- 2011-10-06
- Location
- Glendora, CA