Discovery NM 630, dual detector free-geometry integrated nuclear imaging system. Product Usage: The GE Discovery NM 630
FDA device recall Z-1791-2013 · posted 2013-07-24 · Terminated
Recall details
- Recalling firm
- GE Healthcare, LLC
- Reason for recall
- GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
- Action taken
- GE sent an Urgent Medical Device Correction letter dated June 17, 2013 to all affected customers. The letter identified the affected product, recommended that qualified service personnel maintain the equipment and that Preventative Maintenance procedures are executed according to labeling. Also the Safety Chapter Sections should be re-reviewed with personnel to ensure proper operation of the equipment. GE notified customers again on July 03, 2013, via Urgent Medical Device Recall letter (including confirmation of delivery for US customers) and follow-up telephone call. Customers were instructed to cease using the affected system until GE Healthcare Field Engineer can complete an inspection of the system and perform any necessary repairs at no cost. A GE Healthcare representative will contact the hospitals to arrange for the inspection.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- KPS
- Quantity affected
- 108
- Distribution
- Worldwide Distribution - US Nationwide including Guam, Puerto Rico and the countries of : Afghanistan, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Ri
- Date initiated
- 2013-06-13
- Date posted
- 2013-07-24
- Date terminated
- 2014-06-19
- Location
- Waukesha, WI