Discovery MR750 3.0T Product Usage: Intended use: The Discovery MR750 3.0Tsystem is a whole body magnetic resonance scan

FDA device recall Z-3025-2017 · posted 2017-08-02 · Terminated

Recall details

Recalling firm
GE Medical Systems, LLC
Reason for recall
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Action taken
Consignees received a written recall notice dated 05/17/2017 notifying them of a correction to all units using an Uninterruptable Power Supply. Consignees were instructed not to unplug the unit and to continue normal use. Units with the Eaton 9130 Uninterruptable Power Supply connected to the MR operator console would receive a correction at no charge. This would involve possibly replacing the power cable. GE Representatives will be contacting consignees to arrange corrections. Any questions can be directed to GE Healthcare Service at 1-800-437-1171 or your local service representative.
Root cause
Mixed-up of materials/components
Status
Terminated
Product code
LNH
Quantity affected
68 units in total
Distribution
US Nationwide in the states of : AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI
Date initiated
2017-05-11
Date posted
2017-08-02
Date terminated
2020-09-09
Location
Waukesha, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T, Optima MR450w 1.5T (K160618)Ge Medical Systems, LLC2016-06-09
Discovery MR750 3.0T (K142361)Ge Medical Systems, LLC2014-12-05
DISCOVERY MR750 3.0T ,DISCOVERY MR450 1.5T, DISCOVERY MR750 3.0T,OPTIMA MR450W 1.5T (K142085)Ge Medical Systems, LLC2014-09-26
DISCOVERY MR750 3.0T, DISCOVERY MR450 1.5T, DISCOVERY MR750W 3.0T, OPTIMA MR450W 1.5T (K132376)Ge Healthcare (Ge Medical Systems, LLC)2013-11-15
SIGNA 3.0T MR750 SYSTEM (K081028)Ge Medical Systems, LLC2008-04-25