Dimension Vista System CHEM 1 CAL, Catalog Number KC110, Lot Number 9DM001 The Dimension Vista CHEM 1 CAL is an in vitro

FDA device recall Z-0864-2010 · posted 2010-02-26 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Siemens Healthcare Diagnostics, Inc. has confirmed high accuracy shifts with Dimension Vista URCA Flex reagent cartridge (various lots) when calibrating with some vials from CHEM 1 calibrator lot 9DM001.
Action taken
An Urgent Field Safety Notice dated August 31, 2009, was issued to all affected customers. Customers were instructed to discontinue the use of Dimension Vista System CHEM 1 Cal lot 9DM001 and discard any remaining product. If customers have any question they can contact Technical Solutions at 1-800-241-0420.
Root cause
Other
Status
Terminated
Product code
JIX
Quantity affected
1474 Cartons
Distribution
Nationwide and Foreign: Australia, Canada, Denmark, and Malaysia.
Date initiated
2009-08-31
Date posted
2010-02-26
Date terminated
2010-12-13
Location
Newark, DE