Dimension Vista(R) Cuvettes The Dimension Vista(R) Cuvettes are for use on the Dimension Vista(R) System. The Dimension

FDA device recall Z-0031-2012 · posted 2011-10-07 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Siemens Healthcare Diagnostics has confirmed a low frequency defect in the molding of the Cuvettes that may result in Cuvette Ring and Cuvette Loader jams. It has been determined that the issue is related to a molding defect with the Cuvette flange or "wing" breaking, resulting in an error condition. Therefore, patient results are not impacted.
Action taken
Siemens Healthcare Diagnostics sent a Urgent Field Safety Notice letter dated August 2011, to all affected customers. The letter identifed the product the problem and the action needed to be taken by customer. Customers were instructed to discard all cuvette cartridges of the identified lots. Customers were also informed that until a replacement lot is available, if an error is encountered, they may contact the Technical Solutions Center for assistance at (800) 441-9245. For replacement product , please call the Siemens Atlanta Customer Service Center at (800) 241-0420.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JJE
Quantity affected
1,786
Distribution
Worldwide Distribution -- USA (nationwide) including the states of AL, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, S
Date initiated
2011-08-08
Date posted
2011-10-07
Date terminated
2013-12-24
Location
Newark, DE