Atlas FDA

Dimension Vista LOCI Reaction Vessel

FDA device recall Z-0980-2010 · posted 2010-03-05 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Some vessels may leak. If leakage occurs, patient results will be flagged and are not reportable.
Action taken
The recalling firm issued an Urgent Field Safety Notice dated 11/2009 to all affected customers. Customers have been told to discard any bags of the affected lot. If the affected lot is the only lot the customer has, customers can examine a vessel from the bag. If the vessel mold cavity identity is B8, customers should discard the bag. Thus customers can use the lot if the cavity mold is anything other than B8. In addition, customers have been told to empty the LOCI vessel holder on all systems and to discard vessels in case any affected vessels have been loaded onto the system. For technical questions or concerns, please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JJE
Quantity affected
408 units
Distribution
The device was shipped to medical facilities in CA, CO, FL, GA, IL, LA, MD, MI, MS, NE, NC, NJ, NY, OH, SC, TN, TX, VA, WA, and WI. The product was also shipped to Belgium, France, Germany, Italy, Net
Date initiated
2009-11-24
Date posted
2010-03-05
Date terminated
2011-09-08
Location
Newark, DE

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
DIMENSION VISTA INTEGRATED SYSTEM (K051087)Dade Behring, Inc.2005-07-11