Dimension Vista 500 Intelligent Lab System or Dimension Vista 1500 Intelligent Lab System: An in vitro diagnostic device

FDA device recall Z-0838-2015 · posted 2014-12-20 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Customer complaints of discrepant flagged and/or un-flagged patient and QC results. Reagent probe performance may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server.
Action taken
An Urgent Medical Device Correction, dated November 2014, was issued on November 4th, 2014 to inform customers that Siemens Healthcare Diagnostics has confirmed that reagent probe performance on Dimension Vista systems may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server. Instructions for replacing an associated reagent probe were provided to resolve the issue. Customers are to retain the letter for their laboratory records and forward it to anyone who has received the affected product. Customers are also to complete and return the Field Correction Effectiveness Check Form. Customer should direct their questions to their Siemens Customer Care Center - Technical Solutions at 1-800-441-9250 or their local Siemens technical support representative.
Root cause
Device Design
Status
Terminated
Product code
JJE
Quantity affected
1,949
Distribution
Worldwide Distribution -- Austria, Australia, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic Korea, Sa
Date initiated
2014-11-05
Date posted
2014-12-20
Date terminated
2017-10-06
Location
Newark, DE