Dimension TSPA (Total Prostate Specific Antigen) Flex Reagent Cartridge. The product is shipped 4 cartridges per carton.

FDA device recall Z-0873-03 · posted 2003-05-29 · Terminated

Recall details

Recalling firm
Dade Behring Inc
Reason for recall
test results may be inaccurate by 0.2 to 0.5 ng/mL
Action taken
The recalling firm telephone each customer on 4/24-28/03 to determine if any customer used the lot (CD4085). If so, the customer was instructed to retest all samples with an alternate lot. All customers were instructed to discard any remaining inventory. As a follow up to the phone calls, the recalling firm issued a recall letter dated 4/25/03 informing the customer of the problem and to discard any remaining product.
Root cause
Other
Status
Terminated
Product code
LTJ
Quantity affected
662 cartons
Distribution
The product was shipped to hospitals and clinic laboratories nationwide. There are eight government/military accounts; one of which is foreign.
Date initiated
2003-04-21
Date posted
2003-05-29
Date terminated
2003-07-31
Location
Newark, DE